Добавил:
Опубликованный материал нарушает ваши авторские права? Сообщите нам.
Вуз: Предмет: Файл:
Joseph I. Goldstein, Dale E. Newbury [et al.]. Scanning Electron Microscopy and X-Ray Microanalysis. (2017). (ISBN 978-1-4939-6674-5). (ISBN 978-1-4939-6676-9). (DOI 10.1007978-1-4939-6676-9).pdf
Скачиваний:
20
Добавлен:
11.04.2023
Размер:
68.34 Mб
Скачать

\334 Chapter 20 · Quantitative Analysis: The SEM/EDS Elemental Microanalysis k-ratio Procedure for Bulk Specimens, Step-by-Step

Counts

Counts

160 000

140 000

120 000

10 000

80 000

60 000

40 000

20 000

0

0.0

20 000

15 000

10 000

5 000

0 2.5

Corning Glass A E0 = 15 keV 100 µm square 50 µm square 20 µm square 5 µm square

2 µm square

1 µm square spot

CorningA_1mu_15kV15nA

CorningA_2mu_15kV15nA

CorningA_5mu_15kV15nA

CorningA_20mu_15kV15nA

CorningA_50mu_15kV15nA

CorningA_100mu_15kV15nA

CorningA_spot_15kV15nA

0.2

0.4

0.6

0.8

1.0

1.2

1.4

1.6

1.8

2.0

 

 

 

 

Photon energy (keV)

 

 

 

 

 

Corning Glass A

 

 

 

 

 

CorningA_1mu_15kV15nA

 

 

E0 = 15 keV

 

 

 

 

 

 

CorningA_2mu_15kV15nA

 

 

 

 

 

 

 

 

CorningA_5mu_15kV15nA

 

 

100 µm square

 

 

 

 

 

 

CorningA_20mu_15kV15nA

 

 

 

 

 

 

 

CorningA_50mu_15kV15nA

 

50 µm square

 

 

 

 

 

 

CorningA_100mu_15kV15nA

 

 

 

 

 

 

 

CorningA_spot_15kV15nA

 

20 µm square

5 µm square

2 µm square

1 µm square spot

2.7

2.9

3.1

3.3

3.5

3.7

3.9

4.1

4.3

Photon energy (keV)

. Fig. 20.16  Corning glass A, showing Na and K migration as a function of dose for scanning beams covering various areas (20 keV, 10 nA)

­values for the glass, including the alkali elements Na and K, whereas the point beam results show reductions in the Na and K concentrations. .Figure 20.17 shows that the measured sodium and potassium concentrations increase to reach the synthesized values as the scanned area dimensions are increased to cover areas above 20 x 20-μm (nominal magnification 5 kX) for the particular dose utilized (15 keV, 1500 nA-s). Thus, while scanning a large homogeneous area obviously concedes the spatial resolution capability of electron-excited X-ray microanalysis, this approach may be the most expedient technique to control and minimize alkali element migration.

20 Materials that can serve as useful standards for sodium include certain crystalline minerals such as albite (NaAlSi3O8) in which the sodium is much more stable under electron bombardment. However, even for albite the use of a stationary high intensity point beam may produce significant migration effects, as shown in .Fig. 20.18 for spectra collected with a stationary point beam as a function of dose (upper)

and at the same dose with a fixed beam and two different sizes of scanned areas (lower). Thus, the use of a scanned area rather than a fixed beam may be necessary when collecting a standard spectrum, even on a crystalline material.

20.6.2\ Materials Subject to Mass Loss During Electron Bombardment—the Marshall-Hall Method

Thin Section Analysis

The X-ray microanalysis of biological and polymeric specimens is made difficult, and sometimes impossible, by several forms of radiation damage that are directly caused by the electron beam. At the beam energies used in the SEM (0.1–

30 keV), it is possible for the kinetic energy of individual beam electrons to break and/or rearrange chemical bonds. The radiation damage can release smaller molecules such as

335

20

20.6 · Beam-Sensitive Specimens

. Table 20.14  DTSA-II quantitative analysis of Corning glass A: Comparison of results with a fixed beam and scanned beam (100 μm

square) (15 keV/15 nA); oxygen by assumed stoichiometry

Element

As-­synthesized mass conc

1500 nA-s (fixed beam) raw mass conc

1500 nA-s (100 -μm2 scan) raw mass conc

O

0.4421

0.4644 ± 0.0009

0.4577 ± 0.0006

 

 

 

 

Na

0.1061

0.0098 ± 0.0003

0.1076 ± 0.0004

Mg

0.0160

0.0186 ± 0.0002

0.0164 ± 0.0001

Al

0.0529

0.0058 ± 0.0001

0.0056 ± 0.0000

Si

0.3111

0.3574 ± 0.0007

0.3239 ± 0.0005

K

0.0238

0.0166 ± 0.0003

0.0257 ± 0.0002

Ca

0.0359

0.0386 ± 0.0003

0.0350 ± 0.0001

Ti

0.00474

0.0057 ± 0.0002

0.0051 ± 0.0001

Mn

0.00775

0.0101 ± 0.0003

0.0082 ± 0.0001

Fe

0.00762

0.0090 ± 0.0003

0.0077 ± 0.0001

Cu

0.00935

0.0108 ± 0.0005

0.0096 ± 0.0003

Sn

0.00150

0.0054 ± 0.0007

0.0045 ± 0.0003

Sb

0.0146

0.0140 ± 0.0007

0.0125 ± 0.0002

Ba

0.0050

0.0042 ± 0.0005

0.0042 ± 0.0002

Raw total

 

0.9718

1.0254

. Fig. 20.17  Results of quantitative analysis of Corning glass A as a function of the size of the area scanned. Nominal magnifications indicated

Concentration (mass)

Corning glass A (E0 = 15 keV; 15 nA)

0.12

 

 

 

 

 

 

 

5 kX

2 kX

1 kX

 

 

 

 

 

 

 

 

0.10

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

10 kX

 

 

0.08

 

 

Na

 

 

 

 

 

 

 

 

 

K

 

 

 

 

 

 

0.06

 

 

Ca

 

 

20 kX

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

0.04

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

50 kX

 

 

 

0.02

Fixed

 

500 kX

100 kX

 

 

 

 

Beam

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

200 kX

 

 

 

 

 

0.00

 

 

 

 

 

 

 

 

 

1e-5

1e-4

1e-3

1e-2

1e-1

1e+0

1e+1

1e+2

1e+3

1e+4

Area bombarded (mm2)

\336 Chapter 20 · Quantitative Analysis: The SEM/EDS Elemental Microanalysis k-ratio Procedure for Bulk Specimens, Step-by-Step

Counts

Counts

7 000

 

 

 

Albite

 

 

 

 

Albite_Point_10s_15kV15nA

 

 

 

 

 

 

 

 

 

Albite_Point_50s_15kV15nA

 

 

 

 

 

 

 

 

 

 

 

6 000

 

 

 

E0 = 15 keV

 

 

 

 

Albite_Point_100s_15kV15nA

 

 

 

 

 

 

 

 

 

 

 

 

 

Fixed beam

 

 

 

 

 

 

5 000

 

 

 

10 s spectra

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

4 000

 

 

 

1st (10 s)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

5th (50 s)

 

 

 

 

 

 

3 000

 

 

 

10th (100 s)

 

 

 

 

 

 

2 000

 

 

 

 

 

 

 

 

 

 

1 000

 

 

 

 

 

 

 

 

 

 

0

0.2

0.4

0.6

0.8

1.0

1.2

1.4

1.6

1.8

2.0

0.0

 

 

 

 

 

Photon energy (keV)

 

 

 

 

140 000

 

 

 

 

 

 

 

 

Albite_100kX_15kV15nA

 

 

 

 

 

 

 

 

 

 

 

 

 

Albite

 

 

 

 

Albite_10kX_15kV15nA

 

120 000

 

 

 

 

 

 

 

Albite_spot_15kV15nA

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

E0 = 15 keV

 

 

 

 

 

 

100 000

 

 

 

100 s spectra

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

80 000

 

 

 

1 µm x 1 µm

 

 

 

 

 

 

 

 

 

10 µm x 10 µm

 

 

 

 

 

 

 

 

 

 

 

 

 

 

60 000

 

 

 

Fixed beam

 

 

 

 

 

 

40 000

 

 

 

 

 

 

 

 

 

 

20 000

 

 

 

 

 

 

 

 

 

 

0

0.2

0.4

0.6

0.8

1.0

1.2

1.4

1.6

1.8

2.0

0.0

Photon energy (keV)

. Fig. 20.18  Albite (NaAlSi3O8); E0 = 15 keV, 15 nA: (upper) effect of increasing dose on the Na peak; (lower) effect of fixed beam versus scanned beam on the Na peak

CO, CO2, and H2O that evaporate into the vacuum, causing substantial mass loss from the interaction volume. At the highest beam currents, typically 10–100 nA, used with a focused beam at a static location, it is also possible to cause highly damaging temperature elevations, which further exacerbate mass loss. Indeed, when analyzing this specimen class it should be assumed that significant mass loss will occur during the measurement at each point of the specimen. If all constituents were lost at the same rate, then simply normalizing

20 the result would compensate for the mass loss that occurs duringtheaccumulationoftheX-rayspectrum.Unfortunately, the matrix constituents (principally carbon compounds and water) can be selectively lost, while the heavy elements of interest in biological microanalysis (e. g., Mg, P, S, K, Ca, Fe, etc.) remain in the bombarded region of the specimen and appear to be present at effectively higher concentration than existed in the original specimen. What is then required of any analytical procedure for biological and polymeric specimens is a mechanism to provide a meaningful analysis under these conditions of a specimen that undergoes continuous change.

Marshall and Hall (1966) and Hall (1968) made the original suggestion that the X-ray continuum could serve as an internal standard to monitor specimen changes. This assumption permitted development of the key procedure for beam-sensi- tive specimens that is used extensively in the biological community and that is also applicable in many types of polymer analysis. This application marks the earliest use of the X-ray continuum as a tool (rather than simply a hindrance) for analysis, and that work forms the basis for the development of the peak-to-local background method applied to challenging geometric forms such as particles and rough surfaces. The technique was initially developed for applications in the high beam current EPMA, but the procedure works well in the

SEM environment.

The Marshall–Hall method (Marshall and Hall 1966) requires that several key conditions:

\1.\ The specimen must be in the form of a thin section, where the condition of “thin” is satisfied when the incident beam penetrates with negligible energy loss. For an analytical beam energy of 10–30 keV, the energy loss

337

20

20.6 · Beam-Sensitive Specimens

passing through a section consisting of carbon approximately 100–200 nm in thickness will be less than

500 eV. This condition permits the beam energy to be treated as a constant, which is critical for the development of the correction formula. Biological specimens are thus usually analyzed in the form of thin sections cut to approximately 100-nm thickness by microtome. Polymers may also be analyzed when similarly prepared as thin sections by microtoming or by ion beam milling. Such a specimen configuration also has a distinct advantage for improving the spatial resolution of the analysis compared to a bulk specimen. The analytical volume in such thin specimens is approximately the cylinder defined by the incident beam diameter and the section thickness, which is at least a factor of 10–100 smaller in linear dimensions than the equivalent bulk specimen case at the same energy, as shown in the polymer etching experiment in the Interaction Volume module.

\2.\ The matrix composition must be dominated by light elements, for example, C, H, N, O, whose contributions will form nearly all of the X-ray continuum and whose concentrations are reasonably well known for the specimen. Elements of analytical interest such as Mg, P, S, Cl, K, Ca, and so on, the concentrations of which are unknown in the specimen, must only be present preferably as trace constituents (<0.01 mass fraction) so that their effect on the X-ray continuum can be neglected. When the concentration rises above the low end of the minor constituent range (e.g., 0.01 to 0.05 mass fraction or more), the analyte contribution to the continuum can no longer be ignored.

\3.\ A standard must be available with a known concentration of the trace/minor analyte of interest and for which the complete composition of low-atomic-number elements is also known and which is stable under electron beam bombardment. Glasses synthesized with low atomic number oxides such as boron oxide are suitable for this role. The closer the low–atomic-number element composition of the standard is to that of the unknown, the more accurate will be the results.

The detailed derivation yields the following general expression for the Marshall–Hall method:

CA

Ich

= c

 

 

 

 

 

AA

 

 

(20.9)

Icm

 

 

 

Z 2

 

 

E

 

 

 

 

 

 

Ci

i

loge 1.166

0

 

 

 

Ai

 

 

 

 

i

 

 

 

 

Ji

\

 

 

 

 

 

 

 

 

 

 

In this equation, Ich is the characteristic intensity of the peak

of interest, for example, S K-L2,3 or Ca K-L2,3, and Icm is the continuum intensity of a continuum window of width E

placed somewhere in the high energy portion of the spectrum, typically above 8 keV, so that absorption effects are negligible and only mass effects are important. Ci is the mass

concentration, Zi is the atomic number, and Ai is the atomic weight. The subscript “A” identifies a specific trace or minor analyte of interest (e.g., Mg, P, S, Cl, Ca, Fe, etc.) in the organic matrix, while the subscript “i” represents all elements in the electron-excited region. E0 is the incident beam energy and J is the mean ionization energy, a function only of atomic number as used in the Bethe continuous energy loss equation

Assumption 2 provides that the quantity (CiZi2/Ai) in Eq. (20.9) for the biological or polymeric specimen to be analyzed is dominated by the low-Z constituents of the matrix. (Some representative values of (CiZi2/Ai) are 3.67 (water), 3.01 (nylon), 3.08 (polycarbonate) and 3.28 (protein with S). Typically the range is between 2.8 and 3.8 for most biological and many polymeric materials.) The unknown contribution of the analyte, CA, to the sum may be neglected when considering the specimen because CA is low when the analytes are trace constituents.

To perform a quantitative analysis, Eq. (20.9) is used in the following manner: A standard for which all elemental concentrations are known and which contains the analyte(s) of interest “A” is prepared as a thin cross section (satisfying assumption 3). This standard is measured under defined beam and spectrometer parameters to yield a characteristic-­ to-continuum ratio, IA/Icm. This measured ratio IA/Icm is set equal to the right side of Eq. (20.9). Since the target being irradiated is a reference standard, the atomic numbers Zi, atomic weights Ai and weight fractions Ci are known for all constituents, and the Ji values can be calculated as needed. The only unknown term is then the constant “c” in Eq. (20.9), which can now be determined by dividing the measured intensity ratio, IA/Icm, by the calculated term. Next, under the same measurement conditions, the characteristic “A” intensity and the continuum intensity at the chosen energy are determined for the specimen location(s). Providing that the low-Z elements that form the matrix of the specimen are similar to the standard, or in the optimum case these concentrations­ are actually known for the specimen (or can be estimated from other information about the actual, localized, material being irradiated by the electrons, and not some bulk property), then this value of “c” determined from the standard can be used to calculate the weight fraction of the analyte, CA, for the specimen.

This basic theme can be extended and several analytes— “A,” “B,” “C,” etc.—can be analyzed simultaneously if a suitable standard or suite of standards containing the analytes is available. The method can be extended to higher concentrations, but the details of this extension are beyond the scope of this book; a full description and derivation can be found in Kitazawa et al. (1983). Commercial computer X-ray analyzer systems may have the Marshall–Hall procedure included in their suite of analysis tools. The Marshall–Hall procedure works well for thin specimens in the “conventional” analytical energy regime (E0 10 keV) of the SEM. The method will not work for specimens where the average atomic number is expected to vary significantly from one analysis point to another, or relative to that of the standard. A bulk specimen where the beam-damaged region is not constrained by the